PULCOM Σ-D
In-process
Post-process
RoHS compliant
Size shift
High Accuracy
High Precision Digital Measuring Head

PULCOM Σ-D

Features

Wide Measuring Range (maximum diameter difference of 40 mm)

Using an optical scale, replacing LVDT, as a sensor allows continuous measuring to diameters of 40 mm difference, without setup.

High Precision Measuring

The high precision, high rigidity and high durability fulcrum mechanism is used to make a wide range of high precision measurements possible.

Achieves Superior Linearity for Measurements of All Ranges

A fulcrum measurement method is used in a best-fit method of automatic compensation of problematic centering and measurement deviations caused by diameter deformations, to achieve superior linearity. 

A/B Phase Digital Transmission

Digital transmission shortens the startup time compared to previous size shift models (1/3 of previous) as well as increasing noise immunity.

Improved Maintainability

Removable connectors provided as standard improve maintainability.

External view

Measuring Heads

E-TD-4162, 4163, 4166, 4168

PULCOM Σ-D E-TD-4162、4163、4166、4168

E-TD-4362, 4363

PULCOM Σ-D E-TD-4362、4363
For outer diameter, For inner diameter

Relay box

Relay box

Specifications

Basic specifications
Product name PULCOM Σ-OD-AR
Model E-TD-4162 E-TD-4163 E-TD-4166 E-TD-4167
Product code 4419532 4419533 4419536 4419537
Repeatability ≤ 0.2 μm/4 σ ≤ 0.4 μm/4 σ ≤ 0.2 μm/4 σ ≤ 0.4 μm/4 σ
Measuring force 0.6 to 3 N
Linearity Wide range linearity (reference value) ≤ 0.05% / Full range of diameter differences (when the diameter difference is 25 mm: ≤ 12.5 μm, when the diameter difference is 40 mm: ≤ 20 μm)
Narrow range linearity (reference value) ≤ 0.5% (≤ 5 μm) / ±500 μm
Measuring range 25 mm 40 mm 25 mm 40 mm
Pre travel (one side) ≥ 6.5 mm ≥ 10.3 mm ≥ 6.5 mm ≥ 10.3 mm
Measuring diameter φ 4 to φ 76.5 φ 4 to φ 84 φ 4 to φ 76.5 φ 4 to φ 84
Workpiece profile Continuous surface
Retract type Air type (double-acting)
Retract amount (one side) ≥ 7.3 mm ≥ 11.0 mm ≥ 7.3 mm ≥ 11.0 mm
Retract mechanism (retract air source) 0.35 to 0.5 MPa (return action)
Standard finger length 82 mm 152 mm 82 mm 152 mm
Waterproof standard IP67 compliant (measuring head, relay box)
Cable thickness and length φ 8 × 2 m
Weight (main unit) Approx. 1.4 kg
No. of outputs 2
Remarks A standard finger length of between 82 mm and 152 mm can be supported.
Basic Specifications
Product name PULCOM Σ-D-ID-AR
Model E-TD-4182 E-TD-4186
Product code 4419562 4419566
Repeatability ≤ 0.2 μm/4 σ ≤ 0.2 μm/4 σ
Measuring force 0.6 to 3 N
Linearity Wide range linearity (reference value) ≤ 0.05% / Full range of diameter differences (when the diameter difference is 25 mm: ≤ 12.5 μm, when the diameter difference is 40 mm: ≤ 20 μm)
Narrow range linearity (reference value) ≤ 0.5% (≤ 5 μm) / ±500 μm
Measuring range 25 mm 25 mm
Pre travel (one side) ≥ 6.5 mm ≥ 6.5 mm
Measuring diameter φ 22 to φ 109 φ 22 to φ 109
Workpiece profile Continuous surface
Retract type Air type (double-acting)
Retract amount (one side) ≥ 7.3 mm ≥ 7.3 mm
Retract mechanism (retract air source) 0.35 to 0.5 MPa (return action)
Standard finger length 82 mm 82 mm
Waterproof standard IP67 compliant (measuring head, relay box)
Cable thickness and length φ 8 × 2 m
No. of outputs 2
Weight (main unit) Approx. 1.4 kg
Basic specifications
Product name PULCOM Σ-D-OD-AR (for in-process use)
Model E-TD-4362 E-TD-4363 E-TD-4366 E-TD-4367
Product code 4310052 4310053 4310056 4310057
Repeatability ≤ 0.4 μm/4 σ
Measuring force 1.4 to 3.2 N
Linearity Wide range linearity (reference value) ≤ 0.05% / Full range of diameter differences (when the diameter difference is 25 mm: ≤ 12.5 μm, when the diameter difference is 40 mm: ≤ 20 μm)
Narrow range linearity (reference value) ≤ 0.5% (≤ 5 μm) / ±500 μm
Measuring range 25 mm 40 mm 25 mm 40 mm
Pre travel (one side) ≥ 6.5 mm ≥ 10.3 mm ≥ 6.5 mm ≥ 10.3 mm
Measuring diameter φ 4 to φ 76.5 φ 4 to φ 84 φ 4 to φ 76.5 φ 4 to φ 84
Workpiece profile Continuous surface
Retract type Air type (double-acting)
Retract amount (one side) ≥ 7.3 mm ≥ 11.0 mm ≥ 7.3 mm ≥ 11.0 mm
Retract mechanism (retract air source) 0.35 to 0.5 MPa (return action)
Standard finger length 82 mm 152 mm 82 mm 152 mm
Waterproof standard IP67 compliant (measuring head, relay box)
Cable thickness and length φ 8 × 2 m
Weight (main unit) Approx. 1.4 kg
No. of outputs 2
Remarks The standard finger length is of the two types, 82 mm and 152 mm. Between 82 mm and 152 mm can be supported.